Authors: | Mégane Moreau, Serge Auger, Pierre Picard and Jean Lacoursière |
---|---|
Themes: | High-Throughput, Urine, Cystine, LDTD-MS/MS |
The rapid quantification of cystine in urine can be useful for the detection of cystinuria, an autosomal recessive disorder (MariaLuisa Cabello-Tomás, 1999). Newborn screening for this disease can be performed by thin layer chromatography (Auray-Blais et al., 2021). Cystinuria is characterized by the formation of cystine kidney stones. Overexcretion of poorly soluble cystines causes accumulation of stones in the kidneys due to precipitation (Chillarón et al. 2010). Previously, the screening test for cystine in urine was qualitative . However, rapid quantification of cystine in urine can be helpful in confirming the diagnosis. Additionally, a high concentration of cystine in the urine can lead to kidney stones (obstruction) or kidney failure (Wear et al., 2005).
Our goal for this application note is to develop a diagnostic tool to rapidly quantify cystine in urine, which will allow the rapid diagnosis of certain autosomal disorders, such as cystinuria.
LDTD-MS/MS offers specificity combined with an ultra-fast analysis for an unrivaled quantification method. To develop this application, we focused on performing an efficient sample preparation, which will increase the speed of the analysis process and confirm diagnoses more quickly.
Stock solutions of Cystine (2000 µg/mL) and Cystine-d6 (100 µg/mL) were prepared in sodium hydroxide solution (50 mM). Then, artificial urine samples were spiked to generate a calibration curve and QCs.
LDTD
Model: Luxon S-960, Phytronix
Carrier gas: 6.0 L/min (air)
Laser pattern:
3-second ramp to 55% power
MS/MS
MS model: QTrap® System 5500, Sciex
Curtain: 20
CAD: 8
IonSpray Voltage: 5500
Temperature, GS1 and GS2 set to zero.
Scan Time: 100 msec
Total run time: 0.15 minutes per sample
Analysis Method: Positive MRM mode
Table 1 – MRM transitions for LDTD-MS/MS
Transition | CE (V) | |
---|---|---|
Cystine | 353.0 → 130.0 353.0 → 208.0 |
20 |
Cystine-d6 | 359.0 → 131.0 359.0 → 211.0 |
20 |
Calibration curves ranging from 1 to 100 mg/mL were prepared in synthetic urine. Two sets of QCs were prepared. One set in synthetic urine (QC-L, QC-M and QC-H) and a second set in standard urine (UQC-L and UQC-M). Replicate extractions were deposited onto a LazWell™ plate and dried before analysis. The peak area against the internal standard (IS) ratio was used to normalize the signal.
The calibration curves were plotted using the peak area ratio and the nominal concentration of standards. For the linearity test, the following acceptance criteria was used:
Table 2 shows the inter-day correlation coefficients for Cystine. Values greater than 0.995 are obtained. Figure 3 shows a typical calibration curve result for Cystine.
Table 2 – Inter-day calibration curve correlation coefficients
Cystine | R | |
---|---|---|
Curve 1 | Y = 0.07186 x – 0.00555 | 0.99921 |
Curve 2 | Y = 0.07124 x – 0.01314 | 0.99812 |
Curve 3 | Y = 0.06890 x – 0.04453 | 0.99882 |
Curve 4 | Y = 0.06882 x – 0.04509 | 0.99919 |
For the accuracy and precision evaluation, the following acceptance criteria were used:
For the intra and inter-run precision and accuracy experiment, each QC was analyzed in sextuplicate, on four different runs. Table 3 and 4 show the intra and inter-run precision and accuracy results for Cystine. The obtained %CV was below 15% and the accuracy was within 15% of the nominal value.
Table 3 – Intra-Run Precision and Accuracy of Cystine
Cystine | QC-L1 | QC-M1 | QC-H1 | UQC-L2 | UQC-M2 |
---|---|---|---|---|---|
Conc (mg/mL) | 3 | 50 | 70 | 11.6 | 61.6 |
N (μg/mL) | 6 | 6 | 6 | 6 | 6 |
Mean (μg/mL) | 2.9 | 50.4 | 73.8 | 11.4 | 61.1 |
%CV | 5.4 | 3.2 | 4.4 | 6.6 | 6.3 |
%Nom | 98.3 | 100.8 | 105.5 | 98.1 | 99.2 |
Table 4 – Inter-Run Precision and Accuracy of Cystine
Cystine | QC-L1 | QC-M1 | QC-H1 | UQC-L2 | UQC-M2 |
---|---|---|---|---|---|
Conc (mg/mL) | 3 | 50 | 70 | 11.6 | 61.6 |
N (μg/mL) | 24 | 24 | 24 | 24 | 24 |
Mean (μg/mL) | 3.0 | 50.0 | 69.3 | 11.6 | 61.9 |
%CV | 7.1 | 4.0 | 8.8 | 7.8 | 6.7 |
%Nom | 99.3 | 100.0 | 98.9 | 100.4 | 100.4 |
Following the extraction, sample extracts are kept at 4°C in closed containers protected from light. After a week, sample extracts are spotted on a LazWell™ plate, dried and analyzed. The precision and accuracy results of QCs in synthetic urine samples are reported in Table 5. Results of QCs in urine samples are reported in Table 6. All the results are within the acceptable criteria range for one week at 4°C.
Extracted samples are spotted onto a LazWell™ plate, dried and kept at room temperature for 4 hours before analysis. The precision and accuracy results of QCs in synthetic urine samples are reported in Table 5. Results of QCs in urine samples are reported in Table 6. All the results are within the acceptable criteria range for 4 hours at room temperature.
Table 5 – Wet and Dry Stability of Cystine in Synthetic Urine Sample
Parameters | Dry stability (4 hours / RT) |
Wet stability (1 week / 4°C) |
||||
---|---|---|---|---|---|---|
QC | QC-L | QC-M | QC-H | QC-L | QC-M | QC-H |
Conc. (mg/mL) | 3 | 50 | 70 | 3 | 50 | 70 |
N | 6 | 6 | 6 | 6 | 6 | 6 |
Mean (mg/mL) | 3.1 | 52.1 | 73.9 | 3 | 52.6 | 72.6 |
%CV | 6.2 | 7.0 | 8.6 | 3.3 | 3.3 | 8.4 |
%Nom | 103.8 | 104.1 | 105.6 | 100.37 | 105.2 | 103.7 |
Table 6 -Wet and Dry Stability of Cystine in Urine Sample
Parameters | Dry stability (4 hours / RT) |
Wet stability (1 week / 4°C) |
||
---|---|---|---|---|
QC | UQC-L | UQC-M | UQC-L | UQC-M |
Conc. (mg/mL) | 11.6 | 61.6 | 11.6 | 61.6 |
N | 6 | 6 | 6 | 6 |
Mean (mg/mL) | 11.4 | 58.7 | 11.4 | 62.5 |
%CV | 4.7 | 7.1 | 2.6 | 5.2 |
%Nom | 95.5 | 95.3 | 98.4 | 101.5 |
Urine samples from real patients were tested with this method to correlate with results obtained by traditional LC-MS/MS. The percentage difference between the values is evaluated. A difference of less than 15% is obtained. The results are reported in Table 7.
Table 7 – Cross validation between LC-MS/MS and LDTD-MS/MS
Cystine | U1 | U2 | U3 | U4 | U5 | U6 | U7 | U8 | U9 | U10 | U11 | U12 |
---|---|---|---|---|---|---|---|---|---|---|---|---|
LC (mg/mL) | 11.3 | 20.9 | 17.2 | 25.9 | 16.2 | 27.1 | 19.1 | 29.3 | 3.0 | 12.3 | 11.2 | 19.6 |
LDTD (mg/mL) | 13.1 | 23.4 | 18.4 | 28.1 | 17.0 | 28 | 19.4 | 32 | 2.5 | 12.1 | 11.3 | 20.9 |
%Diff (%) | 7.5 | 5.6 | 3.5 | 4.1 | 2.4 | 1.6 | 0.8 | 4.4 | -8.6 | -0.8 | 0.3 | 3.2 |
The Luxon Ion Source® combined with the Sciex QTrap® 5500 mass spectrometer system enables the rapid analysis of cystine in urine. This method of analysis can thus be applied at the clinical level in order to diagnose certain autosomal recessive disorders such as cystinuria.