Gabapentin Analysis in Oral Fluids

Authors: Mégane Moreau, Serge Auger, Pierre Picard and Jean Lacoursière
Themes: High-Throughput, Gabapentin, Oral Fluid, LDTD-MS/MS
Télécharger le PDF

Introduction

Gabapentin, a drug originally prescribed for the treatment of seizures, is now frequently used off-label for pain management. However, its use can cause side effects such as drowsiness and dizziness, which may impair driving and pose significant safety risks. Considering these issues, the National Safety Council’s Alcohol, Drugs, and Impairment Division (NSC-ADID) has recently upgraded gabapentin’s classification from a Tier II to a Tier I substance, adding it to the standard routine drug testing panel. For this purpose, NSC-ADID established a cutoff concentration of 50 ng/mL in oral fluid.

Our goal for this application note is to use a simple sample preparation method and a fast analysis technique for gabapentin analysis in oral fluid (OF) sample.

The LDTD-MS/MS system using Axino Ion Source offers specificity combined with an ultra-fast analysis for an unrivaled analysis method. To develop this application, we focused on performing a simple sample preparation.

 

Sample Preparation Method

Sample Collection

Oral fluid sample collection is performed with the Quantisal device, which is FDA 510(k) cleared for the collection of oral fluid for drug analysis.

Automated Sample Extraction (SALLE)

The calibration curve was prepared in negative oral fluid samples.

 

LDTD®-MS/MS Parameters

LDTD

Model: Axino, Phytronix

Carrier gas: High

Laser pattern: 6-second ramp to 65% power

MS/MS

MS model: Q-Trap System® 5500, Sciex

Ionization: APCI

Analysis Method: Positive MRM mode

CUR: 20

CAD: 8

 

Table 1 – Positive MRM transitions for LDTD-MS/MS
Drugs Transition CE
Gabapentin 172.0 → 137.0 25
Gabapentin-d10 182.0 → 147.0 25

 

Results and Discussion

Validation Test

To evaluate method precision and accuracy, a three-point calibration curve was employed (1 x cutoff, 2 x cutoff, and 5 x cutoff). For the screening approach, a single-point calibration at the cutoff level was used, applying a linear regression constrained to pass through the origin. The peak area against the internal standard (IS) ratio was used to normalize the signal.

Linearity

For inter-run results, linearity was evaluated. Correlation coefficient (r) must be higher than 0.97. Linearity obtained was higher than 0.991.

Shows the linearity curve of Gabapentin
Figure 1 – Three-point calibration curve for Gabapentin.

Inter-run Precision and Accuracy

Spiked samples around the decision point (Cal-1X, two times cut-off: Cal-2X, and five times cut-off: Cal-5X) and blank solutions are used to validate the precision of the method. For quality control samples, these samples are spiked in real negative matrices. For QC-0.5X, 50% cut-off are spike in sample, and for QC-2X, 200% cut-off are spiked in sample. Replicate extractions are deposited on a Domino LazWell™ plate and dried before analysis.

For the inter-run precision experiment, each fortified sample set is analyzed in triplicate. Table 2 shows the inter-run precision and accuracy results for gabapentin. %CV and %Bias were below 20%.

 

Table 2 – Inter-Run Precision/Accuracy
Gabapentin Cal-1X Cal-2X Cal-5X
Conc (ng/mL) 50 100 250
N 6 6 6
Mean (ng/mL) 49.2 102.0 248.7
%CV 9.7 7.5 5.1
%Bias 1.5 -2.0 0.5

Run Acceptance Criteria for Intra-run

Intra-run validation tests were carried out using a one-point calibration curve and linear through zero regression. Cutoff and Qcs were analyzed in triplicate. Table 3 shows intra-run results.

Table 3 – Intra-Run Precision/Accuracy
Gabapentin Run 1 Run 2
Blank Negative Negative
Conc cutoff (ng/mL) 50 50
N 3 3
Mean (ng/mL) 48.8 50.0
%CV 6.7 6.4
QC-0.5X Negative (3/3) Negative (3/3)
QC-2X Positive (3/3) Positive (3/3)

Multi-Matrix Study

Real OF samples (N=4) have been tested with this method. These samples were readily known to be exempt from gabapentin. All samples were determined to be negative by Axino-MS/MS analysis. Results are reported in Table 4

 

Table 4 – Multi-Matrix Results
Axino Real
M1 Negative Negative
M2 Negative Negative
M3 Negative Negative
M4 Negative Negative

Conclusion

Axino Ion Source® combined to a Sciex Q-Trap 5500 mass spectrometer system allows ultra-fast (10 seconds per sample) for screening gabapentin in oral fluid using a simple and efficient sample preparation method.